Type II Water Purification System F-WPS251
Home Water Purification Systems Water Purification System Type II Water Purification System F-WPS251
Type II Water Purification System F-WPS251
Type II Water Purification System F-WPS251 is a 30 L capacity revolutionary, modular design which helps in elimination of large amount of salts and other minerals from water using 2 stages of RO. It provides Type II level water with consistency and reliability with a TOC level of < 20 ppb. Remove approximately 98% of inorganic ions and 99% of all dissolved organic substances. Equipped with easy replaceable filters. It’s designed to deliver compliant purified water with clear quality parameters and care free maintenance
| Process | Double Pass RO system |
| Output of Pure water | ≥ 10 L / hr. |
| Water Type | Type II |
| Water Tank capacity | 30 L |
| Configuration | Benchtop |
| Output of Ultrapure water | ≥ 1- 1.5 L / min |
| Resistivity at 25 °C | 10-16 M Ω. cm |
| Conductivity at 25 °C | ≤ 5 µs / cm |
| TOC | < 20 ppb |
| Absorbance ( 254 nm ) | ≤ 0.001 |
| Reactive Silica ( SiO2 ) | < 0.01 ppm |
| Heavy metal | < 0.01 ppm |
| Power Supply | 220 V / 50 Hz |
| Power | ≥ 150 W |
| Dimensions ( L x W x H ) | 340 x 550 x 530 mm |
| Weight | 40 kg |
Advanced Double Pass reverse osmosis system for excellent water quality |
Equipped with 30 L water holding tank |
TOC level < 20 ppb |
Type II level water produced |
Source of inlet water – Tap water |
Benchtop model |
LCD display to monitor the intake of water pressure and various parameters intuitively |
Anti-fouling RO membrane prevention programme |
Online monitoring of water resistance |
Ease of cleansing , disinfecting and draining waste water |
Micro filter membrane to filter out pollutants |
Automatic RO membrane antiscale flushing to extend the life of the RO membrane |
Type II Water Purification System is used for conventional physical and chemical analysis experiment. e.g.: Toxicity test,quantitative analysis of trace element, buffer solution.
IC/ICP-MS element analysis